9650-0804-01 Rev. A
ZOLL Propaq M Service Manual
5
Chapter 1
Maintenance Tests
Propaq M Overview
This chapter includes step-by-step procedures for the following types of maintenance tests:
•
Calibration and Calibration Checks
•
NIBP Valve, Pump, and Tubing Integrity Tests
•
Operational Tests
•
Functional Verification Tests
At a minimum, a trained biomedical technician must perform a CO
2
calibration and an NIBP
calibration check once per year. A CO
2
calibration check may be performed at any time to
verify the CO
2
calibration.
The Troubleshooting chapter in this manual may recommend performing one of more of the
NIBP valve, pump, and tubing integrity tests to help identify and correct potential issues with
the NIBP valve, pump, and/or tubing.
It is necessary to perform Functional Verification Tests after a module is repaired to ensure that
the functions of the Propaq M unit work properly.
Because the Propaq M units must be maintained ready for immediate use, it is important for
users to conduct the Operator’s Shift Checklist procedure at the beginning of every shift. This
procedure can be completed in a few minutes and require
s
no additional test equipment. (See
the ZOLL Propaq M Operator’s Guide for the Operator’s Shift Checklist.) It is highly
recommended that the Operational Tests be performed at regular intervals to augment the
automatic testing the Propaq M does to ensure readiness.
Summary of Contents for Propaq M
Page 1: ...Propaq M Service Manual 9650 0804 01 Rev A December 2010...
Page 36: ...CHAPTER 1 MAINTENANCE TESTS 32 9650 0804 01 Rev A...
Page 44: ...CHAPTER 2 TROUBLESHOOTING 40 9650 0804 01 Rev A...
Page 50: ...46 9650 0804 01 Rev A 11 Disconnect the NIBP tube from the rear case NIBP hoses NIBP tube...
Page 52: ...48 9650 0804 01 Rev A 6 Remove the ZIF Zero Insertion Force flex cable using an orange stick...
Page 66: ...CHAPTER 5 REPLACEMENT PARTS 62 9650 0804 01 Rev A...
Page 70: ...CHAPTER 6 FUNCTIONAL DESCRIPTION 66 9650 0804 01 Rev A...
Page 72: ...68 9650 0804 01 Rev A Propaq M Electrical Hardware Interconnect Diagram...
Page 74: ...70 9650 0804 01 Rev A...
Page 78: ...74 9650 0804 01 Rev A...