Propaq M Error Messages
9650-0804-01 Rev. A
ZOLL Propaq M Service Manual
35
TEMP PROBE2 SHORTED
An A/D value was detected
that indicated the probe was
shorted.
Replace the temperature
probe or adapter cable. If
problem persists, replace
the right side panel.
TEMP1 CIRCUIT FAILURE
The software detected an
out-of-range A/D converter.
Remove probe, power
cycle. If problem persists,
replace main board.
TEMP2 CIRCUIT FAILURE
The software detected an
out-of-range A/D converter.
Remove probe, power
cycle. If problem persists,
replace main board.
IBP DEVICE FAILURE
IBP software detected a
problem with the IBP data
that is most likely a hardware
issue.
Replace the main board.
IBP PROBE1 SHORTED
An A/D value was detected
that indicated the probe was
shorted.
Replace the IBP probe. If
problem persists, replace
the right side panel.
IBP PROBE2 SHORTED
An A/D value was detected
that indicated the probe was
shorted.
Replace the IBP probe. If
problem persists, replace
the right side panel.
IBP PROBE3 SHORTED
An A/D value was detected
that indicated the probe was
shorted.
Replace the IBP probe. If
problem persists, replace
the right side panel.
IBP PROBE1 CIRCUIT
FAILURE
An unsupported probe was
inserted or there is a circuit
problem.
Unplug probe, power cycle.
If problem persists, replace
main board.
IBP PROBE2 CIRCUIT
FAILURE
An unsupported probe was
inserted or there is a circuit
problem.
Unplug probe, power cycle.
If problem persists, replace
main board.
IBP PROBE3 CIRCUIT
FAILURE
An unsupported probe was
inserted or there is a circuit
problem.
Unplug probe, power cycle.
If problem persists, replace
main board.
SPO2 DEVICE FAILURE
Communication failure or
device malfunction.
Check SpO
2
communication cables. If
problem persists, replace
the SpO
2
module.
Error Message
Explanation
Technical Action
Summary of Contents for Propaq M
Page 1: ...Propaq M Service Manual 9650 0804 01 Rev A December 2010...
Page 36: ...CHAPTER 1 MAINTENANCE TESTS 32 9650 0804 01 Rev A...
Page 44: ...CHAPTER 2 TROUBLESHOOTING 40 9650 0804 01 Rev A...
Page 50: ...46 9650 0804 01 Rev A 11 Disconnect the NIBP tube from the rear case NIBP hoses NIBP tube...
Page 52: ...48 9650 0804 01 Rev A 6 Remove the ZIF Zero Insertion Force flex cable using an orange stick...
Page 66: ...CHAPTER 5 REPLACEMENT PARTS 62 9650 0804 01 Rev A...
Page 70: ...CHAPTER 6 FUNCTIONAL DESCRIPTION 66 9650 0804 01 Rev A...
Page 72: ...68 9650 0804 01 Rev A Propaq M Electrical Hardware Interconnect Diagram...
Page 74: ...70 9650 0804 01 Rev A...
Page 78: ...74 9650 0804 01 Rev A...