Cleaning and Maintenance
Preventive Maintenance
User Manual
55
►
How to replace the external chiller filter:
Turn the control unit off prior to replacing the filter
Preventive Maintenance
As a rule, replace the chiller filter every six months. However, if the cooling
cue (blue thermometer) is displayed indefinitely, or if the Z802-322 Chiller
Pump Error message is displayed, replace the chiller filter promptly.
►
To replace the external chiller filter:
1.
Slide the filter out of the slot.
2.
Pull the filter downward to slip it out from under the spring.
3.
Discard the used filter.
4.
Remove a new filter from the package.
5.
Slide the bottom of the new filter into the slot.
6.
Fold the front of the filter up under the spring.
7.
Ensure that the front of the filter is fitted smoothly against the frame.
►
How to order:
The replacement part number is FRU-CTU-BAM-103 and can be ordered by contacting your local Allergan