Treatment
Perform a Treatment
User Manual
43
►
To apply a vacuum applicator:
The use of this device on areas with superficially located nerve
branches, arteries, or veins has not been demonstrated to be safe and effective.
Such use may result in injury to the patient.
If the gelpad slips and the cooling surfaces of the applicator come
into direct contact with the patient
’
s skin, tissue injury may result. Inspect the
gelpad and applicator to ensure that the gelpad extends beyond the edges of the
applicator cup.
Use the default vacuum settings or the lowest settings that result in
acceptable tissue draw into the applicator cup.
1.
Press the Vacuum On/Off button on the applicator touch pad.
The vacuum is activated.
The Vacuum On button and the Tissue Draw indicator are displayed.
The Vacuum Status light on the applicator touch pad shines blue.
2.
Place the applicator over the center of the gelpad on the treatment site.
3.
Ensure that the gelpad extends beyond the edges of the applicator cup.
4.
For best results, ensure that tissue is drawn into the applicator cup.
5.
(Optional: Test Vacuum Pressure for Massage)
6.
When the system detects that the applicator is connected to the treatment site, the Start button is
displayed.
The Treatment Status light on the applicator touch pad flashes blue.
Press the Start button.
The Treatment Status light on the applicator touch pad shines blue.
Surface Applicator Treatment
The
CONFIRM?
Site Preparation button is displayed.
Inspect the treatment site to ensure that the skin is intact. Treat
over intact skin only.