User Manual
79
Contents
•
Essential Performance .................................................................... 79
•
Disposal of Hazardous Materials .................................................... 79
•
Environmental Requirements ......................................................... 79
•
Electrical Specifications .................................................................. 80
•
Medical Safety Standards ............................................................... 80
•
Electromagnetic Compatibility ....................................................... 80
•
Data Modem Specifications ............................................................ 84
This product may contain remanufactured parts or parts that have had incidental use, all of which are
equivalent in performance to new parts.
Essential Performance
When cooling to a target temperature that is below 5°C, the device allows cooling to no more than 1°C below
the target temperature. When warming to a target temperature that is above 30°C, the device allows warming
to no more than 1°C above the target temperature. Under steady state conditions, the device controls vacuum
pressure to within ± 1 inches of Hg.
Disposal of Hazardous Materials
Various components of the system may contain materials whose disposal is subject to regulation. The upper
module of the system contains a lithium battery, which is not serviceable by the customer. Dispose of all
components of the system in accordance with applicable regulations. Contact your local environmental control
agency for additional information on recycling or disposing of the system in your area.
Environmental Requirements
The system and its components are designed to operate normally when stored, shipped, and operated under
the following conditions.
Use of the system in an oxygen-rich environment may cause fire. Do
not use the system in an oxygen-rich environment.
CAUTION: The system may not operate as expected if it is stored or operated in
conditions of excessive heat, humidity, or atmospheric pressure. Operate and
store the system in a room that meets the stated requirements.
Shipping / Storage
Operating
Temperature
32°F to 140°F (0°C - 60°C)
59°F to 82°F (15°C - 28°C)
Humidity
10% to 95% (non-condensing)
10% to 70% (non-condensing)
Atmospheric Pressure
Sea level to 10,000 feet at standard pressure
Table 25:
Shipping, Storage, and Operating Requirements
A