Electrical Specifications
System Specifications
80
BRZ-101-TUM-EN2-L
Dimensions of the Control Unit and Modules
Height
Depth
Width
Weight
Control unit alone
47.5 in
120.7 cm
35 in
88.9 cm
24 in
61 cm
215 lbs
97.5 kg
Control unit with support arm
62 in
157.5 cm
n/a
n/a
216 lbs
98.0 kg
Upper module
17 in
43.2 cm
27.25 in
69.2 cm
21.25 in
54 cm
65 lbs
29.5 kg
Base module
30.5 in
77.5 cm
28.5 in
72.4 cm
24 in
61 cm
150 lbs
68.0 kg
Table 26:
Control Unit - Dimensions
Electrical Specifications
Electrical Safety
Class I Equipment, single phase AC, Continuous Operation
Contains Type BF Patient-applied Parts
Water Ingress Protection: Ordinary Equipment, IPX0
REF
Voltage
Frequency
Current
BRZ-CG1-BAM-100
100VAC
50-60 Hz
12A
BRZ-CG1-BAM-110
110-120VAC
50-60 Hz
12A
BRZ-CG1-BAM-220
220-240VAC
50-60 Hz
7A
Table 27:
Electrical Specifications
Fuses
The system contains two internal fuses: Type 3AB (ceramic cartridge), Rating: 250VAC, 6.25A, Slo-Blo. The fuses
are not serviceable by the customer.
Medical Safety Standards
The system complies with the following medical safety standards:
•
IEC 60601-1: 1998 + A1, A2
•
IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007)
•
EN 60601-1: 2006
•
CAN/CSA C22.2 No 60601.1: 08
•
ANSI/AAMI ES 60601-1: 2005 / AS: 2010
•
AS 3200.1.0: 1998 + A1/NZS 6150: 1990 + A1
•
Electromagnetic Compatibility (EMC) EN 60601-1-2: 2007
Electromagnetic Compatibility
The system has been tested and found to comply with Medical Standard Electromagnetic Compatibility (EMC)
EN 60601-1-2: 2007. The system complies with the standards outlined below.
This system requires special precautions to ensure electromagnetic compatibility with other electrical medical
devices. To ensure EMC, the system must be installed and operated according to the information provided in
this manual.