Perform Another Treatment
Treatment
46
BRZ-101-TUM-EN2-L
3.
Press the Next button.
4.
Press the appropriate button for the current patient and treatment site.
The Profile panel is displayed.
•
See Select a Profile on page 41.
►
To perform a treatment on the next patient:
CAUTION: When the vacuum is turned off or the securement system straps are
released, the applicator may disengage from the patient. The applicator could
fall and be damaged or cause injury. Grasp the head of the applicator firmly
before turning off the vacuum or releasing the securement system straps.
Remove a vacuum applicator:
Remove a surface applicator:
Grasp the applicator and turn off the vacuum.
Grasp the applicator and release the securement system
straps.
Remove the applicator from the patient.
Remove the applicator from the patient.
Place the applicator head on top of the control unit with
the cooling surfaces facing downward.
Place the applicator head on top of the control unit with
the cooling surfaces facing upward.
Allow gel to drain onto a towel or other absorbent
material.
n/a
Remove the gelpad or gel from the treatment site.
Remove the liner, gelpad, and foam borders from the
treatment site.
Remove the liner from the applicator cup.
n/a
Discard the used gelpad or gel, and liner according to your
site
’
s medical waste protocols.
Discard the used liner, gelpad, foam borders, and
securement system according to your site
’
s medical waste
protocols.
1.
If the
Same Patient
and
Next Patient
buttons are displayed, press the
Next Patient
button.
If the Profile panel is displayed, select a profile.
•
See Enter Patient Data on page 40.
Expired Card
If the card is expired, a recoverable exception is displayed.
1.
Remove the card from the applicator.
2.
Press the Next button to clear the message.
3.
Insert a new card into the slot on the applicator.
The system authenticates the card.
4.
When authentication is complete, press the Next button.
The profile screen is displayed.