Perform Another Treatment
►
To perform another treatment on the same patient:
CAUTION: When the vacuum is turned off or the securement system straps are
released, the applicator may disengage from the patient. The applicator could
fall and be damaged or cause injury. Grasp the head of the applicator firmly
before turning off the vacuum or releasing the securement system straps.
Remove a vacuum applicator:
Remove a surface applicator:
Grasp the applicator and turn off the vacuum.
Grasp the applicator and release the securement system
straps.
Remove the applicator from the patient.
Remove the applicator from the patient.
Place the applicator head on top of the control unit with
the cooling surfaces facing downward.
Place the applicator head on top of the control unit with
the cooling surfaces facing upward.
Allow gel to drain onto a towel or other absorbent
material.
n/a
Remove the gelpad or gel from the treatment site.
Remove the liner, gelpad, and foam borders from the
treatment site.
Discard the used gelpad or gel according to your site
’
s
medical waste protocols.
Discard the used liner, gelpad, and foam borders according
to your site
’
s medical waste protocols.
The Same Patient
and Next Patient buttons are displayed.
(If the card is expired, see Expired Card on page 46.)
1.
Press the
Same Patient
button.
2.
On the Body Profile screen, select a treatment site.
In this example, the lower abdomen is selected for a female patient.
Refer to the Preface for warnings and cleared intended use.