130
Preparations
G-30-1530-en
OPMI® VISU 210 on S8, S81, S88 Suspension Systems
Issue 1.0
Printed on 19. 04. 2004
•
Screw in the securing screw (7) and tighten it firmly . The securing
screw (7) must go into the groove (9). This is ensured when the secu-
ring screw is flush with the outer surface of the coupling.
•
Insert the coupling (12) including the surgical microscope from above
into the receptacle (11) on the suspension arm and tilt the coupling
downward into its vertical position.
•
Firmly tighten the mounting screw (3) using a 5 mm Allen key.
•
Plug the cable clip (10) into the opening (13) of the coupling.
•
Then perform the balance setting procedure.
Warning!
Before using and after re-equipping the unit, always make sure that se-
curing screws (3), (4) and (7) have been tightened firmly.
Note:
When mounting surgical microscopes with an integrated coupling, steps
3 to 9 need not be performed.
MISU241