128
Preparations
G-30-1530-en
OPMI® VISU 210 on S8, S81, S88 Suspension Systems
Issue 1.0
Printed on 19. 04. 2004
Attaching the equipment
Mounting the surgical microscope
Warning!
The maximum weight of the microscope including accessories must not
exceed 20 kg!
•
Leave the suspension arm locked in its horizontal position until you
–
have mounted and secured the complete equipment,
–
and made the electrical connections.
•
Then perform the balance setting procedure.
•
Turn off the system at the power switch.
•
Bring the suspension arm into its horizontal position, pull out locking
knob (1) and turn it clockwise or counterclockwise through 180°. At the
same time, slightly move the suspension arm up and down until the
lock snaps in. This prevents the suspension arm from uncontrollably
moving upward when insufficient weight is attached.
•
Use a 4 mm hex key to loosen mounting screw(3) by a few turns.
•
Tilt coupling (2) upward and remove it in the upward direction.
•
Loosen locking screw (6) by a few turns.
•
Give securing screw(7) a few turns to loosen it.
•
Slightly lubricate microscope shaft (8) (e.g. with instrument grease or
vaseline).
•
Slide coupling (2) from above over microscope shaft (8). Screw in
mounting screw (5) from above and firmly tighten securing screw (4)
using a 4 mm hex key.
MISU241