Two different illumination systems are available for the suspension sy-
stem.
1
Xenon illumination system
The xenon illuminator is equipped with an illumination system for fiber
illumination. The xenon lamp generates light whose spectrum resem-
bles that of natural daylight. Regardless of the brightness setting, the
color temperature of the light always remains the same. Normal day-
light film without any additional conversion filters can therefore be
used for photographic documentation. The lamp module contains two
xenon lamps. The second lamp is used as a backup lamp which has
to be manually swung into the illumination beam path if the first lamp
fails. You have to pull out the lamp module all the way before being
able to swing in the backup lamp.
Warning!
When using xenon illumination, only operate the system with special
xenon lamps approved by Carl Zeiss. If any other than Carl Zeiss-ap-
proved xenon lamps are used, there is the risk of severe injury to the pa-
tient's eye.
2
Halogen illumination
The halogen illuminator is equipped with an illumination system for
fiber illumination. The lamp housing contains a backup lamp which is
automatically swung into the illumination beam path when the first
lamp fails.
If required, the illumination system can be equipped with a second
lamp housing so that two separate illumination systems are available
for fiber illumination. The second illumination system can be used, for
example, for a fiber slit lamp or a dual fiber illumination system.