46
Description
G-30-1530-en
OPMI® VISU 210 on S8, S81, S88 Suspension Systems
Issue 1.0
Printed on 19. 04. 2004
VISU 210 surgical microscope
Intended use
Note:
The illumination system of the VISU 160 surgical microscope always con-
tains a UV blocking filter.
The use of this filter ensures that the illumination intensity lies below 50
µW/cm
2
in the range between 305 nm and 400 nm.
This helps the surgeon to reduce the risk of phototoxic retinal injury in the
patient.
Description of the modules
The VISU 210 surgical microscope is comprised of the following modules:
1
X-Y coupling
The X-Y coupling allows motorized fine positioning of the surgical mi-
croscope in a horizontal plane. The range of travel is 40 mm x 40 mm.
The speed of travel can be set on the control panel of the suspension
system.
The X-Y coupling is provided with a recentering mechanism. When
you press actuator button (2) or button (4) or (5) of the foot control
panel,
–
the X-Y coupling adopts its center position,
–
the focusing system of the surgical microscope is set to its initial
position (3), if the XY-RES function has been selected in configu-
ration mode 1, see page 185.
–
the zoom system is set to a preselected magnification factor, if the
XYZ-RES function has been selected. This function is only availa-
ble with the S88 floor stand.
4
5
MISU241