G-30-1530-en
OPMI® VISU 210 on S8, S81, S88 Suspension Systems
Issue 1.0
Printed on 19. 04. 2004
Maintenance / Further information
207
Motorized
focusing
and
zoom functions of surgical
microscope are inoperative.
Manual
function
is
acti-
vated. (Yellow LED above
the key lights up).
Switch off manual function.
-
Suspension arm is in hori-
zontal position and cannot
be moved upwards or down-
wards.
Suspension arm still locked. Pull out locking device and
turn though 180°.
-
In connection with three
successive beeps.
–
Error message during
software
check
after
power-on of the suspen-
sion system.
–
Error message in the
case of internal system
error.
Manual operation is not pos-
sible.
Contact service dept.
-
S88 floor stand only: Stand
wobbles.
Floor not level. Stand base
not
appropriately
posi-
tioned.
Slightly turn stand base. Ar-
ticulated arm should be po-
sitioned at a right angle with
the tilt axis.
Problem
Possible cause
Remedy
See
MISU241