G-30-1530-en
OPMI® VISU 210 on S8, S81, S88 Suspension Systems
Issue 1.0
Printed on 19. 04. 2004
Maintenance / Further information
205
Suspension systems
Surgical
field
illumination
too bright.
Brightness
level
set
too
high.
Adjust brightness using the
control on the suspension
system or the foot control
panel.
-
Switch off illumination on
suspension system.
Illuminate
surgical
field
using an OR illuminator.
Contact service dept.
-
Focusing system inopera-
tive.
Focusing system does not
work, or moves to upper or
lower end position.
Use suspension arm to set
correct focal plane.
Contact service dept.
-
Zoom system inoperative.
Zoom
system
does
not
work, or moves to upper or
lower end position.
Manually adjust magnifica-
tion using zoom knob.
Contact service dept.
-
Zoom and focusing systems
inoperative.
Functions not correctly set
on suspension system.
Contact service dept.
-
Microscope motion too stiff.
Friction adjusting knob on
suspension
system
tight-
ened too firmly.
Slightly loosen friction ad-
justing knob.
page 128
DeepView inoperative
DeepView is on mechanical
end stop.
Press the DeepView button
to deactivate the function.
-
Problem
Possible cause
Remedy
See
Problem
Possible cause
Remedy
See
No function at all.
Line power failure.
Contact
in-house
electri-
cian.
-
Power switch of suspension
system not switched on.
Press power switch.
-
Automatic circuit breaker in
power switch of suspension
system has been activated.
Press power switch again.
-
MISU241