Failure of lamp control
•
Press the Manual key (1) if brightness control is no longer possible.
Note:
When the manual function has been activated, all electrical control sys-
tems are inoperative. The lamp brightness is automatically adjusted to a
fixed setting.
Failure of focusing system
•
Press Manual key (1) if, for example, the focusing system always
moves into its upper or lower end position.
•
Focus by moving the suspension arm of the ceiling mount (floor stand)
up or down.