Disinfecting the control keys
To be able to use the unit in ORs, we recommend disinfecting it using
MELISEPTOL disinfectant solution (B. Braun, Melsungen AG). Carl Zeiss
can supply MELISEPTOL and you can obtain it in many countries from
B. Braun representatives.
Caution:
–
Wear disposable plastic gloves to prevent skin contact with the disin-
fectant.
–
MELISEPTOL is inflammable (flame point 31 °C). Please read the
product information from B. Braun, Melsungen AG.
To apply MELISEPTOL, proceed as follows:
•
Switch the unit off before disinfecting.
•
Apply the disinfectant across the entire surface of the control panel.
Do not let any disinfectant seep into the unit.
•
Leave the disinfectant on the unit for approx. 30 minutes.
•
Afterwards wipe the disinfectant off the surface using a sterile, lint-free
cloth.
•
Dispose of the gloves and the cloth as normal waste.
You may obtain the following articles from Carl Zeiss:
Disposable gloves:
Cat. No.
1 l MELISEPTOL in vario-bottle
INR 0103.907
5-l refill bottle
INR 0103.908
MELISEPTOL HBV spray, 500 ml
INR 0103.910
MELISEPTOL HBV cloths
INR 0103.911
size 1 (large)
INR 0117.736
size 2 (medium)
INR 0117.737
size 3 (small)
INR 0117.738