Operation
G-30-1530-en
OPMI® VISU 210 on S8, S81, S88 Suspension Systems
Issue 1.0
Printed on 19. 04. 2004
Checklist
Warning!
If a function fails, you must not use this instrument for safety reasons. Cor-
rect the fault (see the "Troubleshooting table“) or contact our service dept.
Always check the following points before surgery (without patient!):
–
Check that the correct rated voltage has been set.
–
Check that all cables have been connected.
–
Check that the light guide has been connected.
•
Turn on the system at the power switch of the suspension system.
VISU 210 surgical microscope
Check the zoom function
–
Press the appropriate button on the foot control panel.
Check the focusing function
–
Press the appropriate button on the foot control panel.
Friction adjustment of the surgical microscope
–
Check that the friction of the surgical microscope's rotation has been
adjusted as required using the friction adjustment knob on the suspen-
sion system.
Locking of the assistant's microscope
–
Check that the assistant's microscope is locked in the working position
required (on the right or left of the main microscope).
Speeds of the microscope functions
–
Check that the speeds of the microscope functions have been adjust-
ed as required on the suspension system.
Eyepieces / binocular tube
–
Check that the surgical microscope and the tube are in a position con-
venient for you.
–
Check that the correct interpupillary distance has been set.
–
Check that the eyecups have been adjusted in such a way that you
can see the full field of view.
–
Check that the correct prescription has been set on the diopter scale.
–
Check that image quality is the same throughout the entire magnifica-
tion range.