234
Technical data
G-30-1530-en
OPMI® VISU 210 on S8, S81, S88 Suspension Systems
Issue 1.0
Printed on 19. 04. 2004
Electrical data of S88 floor stand
Power connection
Only connect the suspension system to wall out-
lets which are provided with a properly connected
protective ground conductor.
Rated voltage
115 VAC (100...125 VAC± 10%)
230 VAC (220...240 VAC± 10%)
Current
consump-
tion
115 VAC max.10 A
230 VAC max. 8 A
Rated frequency
50...60 Hz
Fuses
Automatic circuit breaker
Electrical
outlets
–
Power outlet 115/230 VAC, max. 5 A
–
power outlet 115/230 VAC, max. 2 A,
–
(via power switch),
–
X-Y coupling
–
Surgical microscope
–
Remote control socket for an external signal
of a maximum of 24 V / 0.5 A.
Electrical
safety
complying with IEC 601-1/EN 60 601-1/ UL 2601-
1; CSA C22.2 No 601.1
Protection class I,
type B equipment
EMC
requirements
complying with EN 60601-1-2: 1993
Class A (hospital)
Remote control sock-
et
1
2
3
1
2
3
MISU241