G-30-1530-en
OPMI® VISU 210 on S8, S81, S88 Suspension Systems
Issue 1.0
Printed on 19. 04. 2004
Safety
23
Before every use and after re-equipping the instrument
•
Make sure that all ”Requirements for operation” are fulfilled.
•
Go through the checklist.
•
Re-attach or close any covers, panels or caps which have been re-
moved or opened.
•
Pay special attention to warning symbols on the instrument (triangular
warning signs with exclamation marks), labels and any parts such as
screws or surfaces painted red.
•
Do not cover any ventilation openings.
For every use of the instrument
General
•
Never operate the system unattended.
•
Avoid looking directly into the light source, e.g. into the microscope ob-
jective lens or a light guide.
•
When the illumination is on, the light guide must be connected at both
ends. Otherwise there is a risk of fire or burn injuries.
•
Make sure that the instrument has been switched off before you
change the xenon lamp module. When switched on, the ignition sys-
tem generates high voltage.
Xenon lamps feature high luminance and a spectrum resembling that of
natural daylight. Therefore, only special xenon lamps approved by Carl
Zeiss must be used in ophthalmology.
•
Any kind of radiation has a detrimental effect on biological tissue.This
also applies to the light illuminating the surgical field. Please therefore
reduce the brightness and duration of illumination on the surgical field
to the absolute minimum required.
•
When operating on the eye, always use a GG 475 protection filter to
ensure that the patient's retina is not exposed to unnecessary (blue)
radiation (retinal injury).
S88 floor stand
•
Using the locking pedal on the base, secure the stand in position.
Make sure that the stand is stable and cannot roll away.
MISU241