14-11
IM CW120-E
Specifications
14
(3) 96033 Clamp-on Current Probe
96033 (Max. 50-A Range)
Item
Measurement range
Output voltage
Accuracy
Temperature coefficient
Maximum allowed current
Output impedance
External magnetic field effects
Conductor position effects
Circuit voltage
Withstanding voltage
Measurable conductor diameter
Operating temperature and
humidity ranges
Storage temperature and
humidity ranges
Applicable locations
Dimensions (W
×
H
×
D)
Weight
Output terminals
Output cable length
Standard accessories
Amplitude
Phase
0 to 50 Arms AC
0 to 0.5 Vrms AC (10 mV/A)
±
1.0% of rdg
±
0.3 mV (20 Hz to 45 Hz)
±
0.5% of rdg
±
0.1 mV (45 Hz to 66 Hz)
±
0.8% of rdg
±
0.2 mV (66 Hz to 1 kHz)
±
1.0% of rdg
±
0.3 mV (1 kHz to 5 kHz)
±
3.0% of rdg
±
0.4 mV (5 kHz to 20 kHz)
±
1.0
°
(for 1 to 50 A, 45 Hz to 1 kHz)
Under the conditions of the temperature at 23
°
C
±
5
°
C, humidity at
35 to 75% RH, with sine wave input
0.05% of full scale/
°
C in ranges of 0
°
C to 50
°
C
130 Arms AC (45 Hz to 1 kHz)
Approximately 18
Ω
0.1 A equivalent or less (at 400 A/m, 50/60 Hz)
±
0.5% (at 1 to 50 A, 45 Hz to 1 kHz)
300 Vrms AC maximum
3.7 kVrms AC for 1 minute (core-to-casing, and core-to-output)
18 mm diameter maximum
0
°
to 50
°
C, 5 to 85% RH (no condensation)
–20
°
to 60
°
C, 90% RH or less (no condensation)
Indoors, altitude of 2000 m or lower
Approximately 52
×
106
×
25 mm
Approximately 220 g
Banana plugs (safety terminals)
Approximately 3 meters
One User’s Manual, ring markers (4 colors
×
2 each, L4007MG)
Safety standards:
Compliant with EN61010-1, EN61010-2-032
• Overvoltage category III (CAT III, max. input
voltage 300 Vrms)
• Pollution degree 2
EMC (emission):
Compliant with EN55011, Group 1, Class A; and
EN61326: 1997
These products are goods in Class A (designed
for industrial environments). For home use, the
user must implement adequate measures to
prevent influences from radio noise emission.
EMC (immunity):
Compliant with EN61326: 1997
Effect on measurement input under immunity
environment:
±20% of range or less
14.2 Specifications of Clamp-on Current Probes