Instantaneous Value Measurement
8
NOTE
• The file name will be given automatically.
• The file number is renewed in the following cases: (1) the data record number has exceeded
99; (2) the wiring method has changed; (3) the
or
actions have been carried out
in the setting screen; (4) the PC card is removed or a PC card is inserted; and (5) the power is
turned off once and then back on.
• Up to 30 files can be saved to a PC card (corresponding to file numbers from 000 to 029);
attempting to save too many files causes
to appear. This will disappear and the
measurement screen will return when any key is pressed. Delete an unnecessary file or files
and try saving again.
• If there is something wrong with the inserted PC card, data cannot be saved and
appears. (The measurement screen will return when any key is pressed.)
• When the PC card becomes full, the
mark appears and data will not be saved.
• Saving the measured data when the communication device is set to a printer, causes the
mark to flash and prints the data at the same time as the data saving.
File Name and Format
Measured data are saved to a CSV format file and the file name (“MWTH”
followed by a number) is given automatically.
File name = MWTHxxx.CSV
Name (xxx): The smallest, unused number from 000 to 029
Filename extension: Always CSV
Command:
INTEgrate:STORe:FILEname
Filename
Example of Saved, Measured Data
Shown below is an example of a measured-data file (that was saved in a PC
card) when opened from a spreadsheet application that has a CSV file
converter. (The field configuration varies depending on the measurement
conditions.)
CW12x
File Type
Measure
Mode
OUTPUT
DATE
2001/11/11
2001/11/11
2001/11/11
2
1
OUTPUT
TIME
10:10:00
10:15:00
10:21:00
INTEG START
DATE
0000/00/00
0000/00/00
0000/00/00
INTEG START
TIME
0:00:00
0:00:00
0:00:00
ELAPSED
TIME
0:00:00
0:00:00
0:00:00
V1
I1
P
Q
PF
F
Wh(+)
Wh(–)
8.3 Saving Data