6-7
IM CW120-E
Setting Ranges and Wiring Method
6
6.3
Changing Current Range Setting
●
Procedure to Change Current Range Setting
(1) The
sign is displayed lowermost on the screen, above the current
range currently selected. To change the range, press the
key when
the measurement screen is displayed.
Current ranges
A RANGE key
A
RANGE
5A
10A 20A 50A 100A 200A 500A 1000A
Each press of the
key changes the range selection sequentially as
follows:
When the 96030 clamp-on probe is selected for use
20 A
50 A
100 A
200 A
Default: 20 A
TIP
• If a change to the range setting is incompatible with the conditions derived from other settings,
is displayed and the change is ineffective. First change other settings such as the current range, VT
ratio, CT ratio, and wiring method, then change the voltage range.
• The voltage range cannot be changed during continuous measurement.
• The voltage range cannot be changed when the electric energy count is not zero. Pressing the
key at this time causes
to appear on the screen (this will disappear and the measurement
screen will return when any key is pressed). Clear the electric energy count first. For details of how to
do this, see Section 9.2.3, “Clearing Electric Energy Count,” or Section 7.2.4, “Clearing Electric Energy
Count and Resetting System,” in Chapter 7, “Making Settings.”