14-5
IM CW120-E
Specifications
14
14.1 Specifications of CW120/121
(11)
External Control Input/Output (provided only with model having
RS-232 interface)
Can be used for remote start and end of integration for electric energy
measurement.
External control input
TTL level or voltage-free contact
External control output
TTL level or contact
(12)
Other Functions
Clock (accuracy: ±100 ppm, typical), key-lock, system resetting
(13) General Specifications
Applicable locations:
Indoors, altitude of 2000 m or lower
Operating temperature and humidity ranges
0°C to 50°C, 5 to 85% RH (no condensation)
Storage temperature and humidity ranges
–20°C to 60°C, 90% RH or less (no condensation)
Insulation resistance
50 M
Ω
or more at 500 V DC
between voltage input terminals (short-circuited)
and casing;
between voltage input terminals (short-circuited)
and current/communication/external control
input/output terminals (short-circuited);
between power line and casing; and
between power line and current/communication/
external control input/output terminals (short-
circuited)
Withstanding voltage
5550 V AC for 1 minute between voltage input
terminals (short-circuited) and casing;
3250 V AC for 1 minute between voltage input
terminals (short-circuited) and current/
communication/external control input/output
terminals (short-circuited); and
2300 V AC for 1 minute between power line and
casing, and between power line and current/
communication/external control input/output
terminals (short-circuited)
Power supply
100 to 240 V AC ±10%, 50/60 Hz
Power consumption
Maximum 8 VA
External magnetic field effects
Within accuracy range at 400 A/m
Dimensions (W
×
H
×
D) Approximately 117
×
161
×
51 mm
Weight
Approximately 0.6 kg
Terminals
Voltage inputs
CW120: 3 jacks for banana plugs (safety
terminals)
CW121: 4 jacks for banana plugs (safety
terminals)