7-20
IM CW120-E
7.2.12 Making Communication Settings (
)
Description
Follow the procedure below to set the baud rate, data bits, parity bit, and stop
bits to match the specifications of the device to connect to.
• Baud rate: 1200, 2400, 4800, 9600, or 19200 bps (9600 bps by default)
• Data bits: 7 or 8 (8 bits by default)
• Parity:
for no parity (default)
for even parity
for odd parity
• Stop bits: 1 or 2 (1 bit by default)
Procedure
(1) Using the
keys in the setting item selection screen, make the
mark and the current communication settings appear on the screen.
Then, press the
key.
The baud rate setting flashes.
DISP
ENTER
MEAS/SET
ESC
MEAS/SET key
Data bits-parity-stop bits
Baud rate
UP/DOWN/LEFT/RIGHT keys
ENTER key
Communication Setting Screen
ESC key
(2) If necessary, change the baud rate using the
keys. The baud rate
changes sequentially as follows.
7.2 Setting Each Item