Normal Setting
Sets and queries the date and time
×
○
Command = DTyyyymmddhhmmss<Delimiter>
→
Answer = DTyyyymmddhhmmss<Delimiter>
Command = DT? (CrLf)
→
Return = yyyyy/mm/dd, hh:mm:ss<Delimiter>
Parameter (default Value)
yyyy: Year (2006) 4 byte, mm: Month (04) 2 byte, dd: Day (01) 2 byte
hh: Hour (00) 2 byte, mm: Minute (00) 2 byte, ss: Second (00) 2 byte
Setting (confirmation) of the 2 leftmost digits of the year is only possible
for the communication function.
(Using the operation keys to change the values of the 2 leftmost digits on
the LCD will not move up or down to the 2 rightmost digits.)
DW
Normal Setting
Decreases the m (th) digit of the source value by 1 digit
○
○
Command = DWm<Delimiter>
→
Answer = DW, OK<Delimiter>
Parameter m = 1 to 5 (1: Least Significant Digit to 5: Most Significant
Digit)
ESC C
or
RC
Normal Setting
Initializes the setting information (setting conditions)
("ESC" = ASCII 0 &batu&1B)
○
×
Command = ESC C<Delimiter
or
Command = RC <Delimiter>
The following settings (common setting item) are not initialized.
• Auto power off setting
• Communication setting
• International temperature standard selection (ITS-90/IPTS-68)
• Time setting
ESC S
Normal Setting
Outputs the status byte
("ESC" = ASCII 0 x 1B)
○
○
Command = ESC S<Delimiter>
The status byte is output. (Decimal number)
(Refer to "8.6 Status Byte Format.")
H
Normal Setting
Sets and queries the output header of the OD and OM commands
○
○
Command = Hm<Delimiter>
→
Answer = Hm <Delimiter>
Command = H?<Delimiter>
→
Return = Hm <Delimiter>
Parameter m=0: No/1: Yes
default Value of m=0 (No)
HD
Normal Setting
Holds and queries the measurement value indication
○
×
Command = HDm<Delimiter>
→
Answer = HDm <Delimiter>
Command = HD?<Delimiter>
→
Return = HDm <Delimiter>
Parameter m=0: Indication Update/1: Indication Hold
8.3 Detailed Description of Commands