9-8
IM CA150E
9.3 Verification after Calibration
After calibration is complete, verify that the instrument was calibrated
correctly and that the calibration values were written to memory.
Verification Procedure
After calibration is complete, turn off the power.
Turn the power back on and verify generation and measurement in
normal mode (not calibration mode).
Use the standard device used for calibration.
(Check the calibration points.)
9.4 Calibration of Temperature Ranges
Calibrating thermocouple (TC) ranges for temperature measurement
involves using special equipment for calibrating the RJ sensor
(reference junction compensation).
9.3 Verification after Calibration