3-8
IM CA150E
3.2 Connecting a Power Supply
• NiMH Battery Life
The battery can be charged approximately 300 times.
(This number varies depending on the operating environment.)
The life of the battery is over when the low battery level indication appears
only a short time after the battery is fully charged. In such a case,
replace the NiMH battery pack with a new one.
For details on disposing, refer to "10.2.2 How to Replace and Dispose the
Batteries."
3.2.4 Fuse
CAUTION
Be sure to use the designated fuse.
A fuse for current input protection is inserted in the side (fuse holder) of the
instrument.
Part number: A1566EF, Rating: 125 mA/250 V FAST
Procedure for Replacing the Fuse
When replacing the fuse, turn the M- part of the fuse holder with a flat-blade
screwdriver and remove the holder. Then, replace the fuse, reinsert the fuse
holder, and turn the fuse holder with the screwdriver.
Fuse holder
Fuse
(A1566EF)