4-24
IM CA150E
4.9 Sweep Output Functions
4.9.3 Program Sweep Function
Multiple setting values saved with the SAVE key can be output in order
as shown in the figure below.
5V
0V
Memory No.
SOURCE
ON
0
1
2
3
4
3
2
1
0
1
2
•••
When the start number was set to 0
Memory Number
Function
Range
Source Value
(Setting Value)
0
1
2
3
4
5
•
•
•
990
DCV
DCV
DCV
DCV
DCV
Unused
•
•
•
•
10 V
10 V
10 V
10 V
10 V
Unused
•
•
•
•
5.0000V
1.0000V
4.0000V
2.0000V
3.0000V
Unused
•
•
•
•
Tip
• This is only valid for the saved data memory information of SOURCE.
(The saved information of measurement conditions is ignored.)
• If there is an unused number during the step operation or a number saved with
different FUNCTION and RANGE conditions, the step operation returns to
the starting number and repeats the step operation.
• The frequency or pulse source values saved in the data memory cannot be used
for program sweep.
You can change the sweep interval time.
The default setting (at shipment) is 5 seconds (FAST).
Set 5 seconds (FAST) or 10 seconds (SLOW) in setting mode.
The interval setting (FAST, SLOW) is also common to step sweep and
linear sweep.
When a sweep function is enabled, a setting mode cannot be selected.
Disable the sweep function by pressing the SWEEP SET key and
switch to the normal mode.