8-6
IM CA150E
CD
Normal Setting
Sets the source value during calibration
×
○
*
Command = CD<Delimiter>
→
Answer = CD <Delimiter>
During generation in calibration mode, sets the current output source
values as the calibration values for the selected function, range, and
scale (+FS/0).
CL
Normal Setting
Sets and queries the calibration item
×
○
*
Command = CLm<Delimiter>
→
Answer = CLm <Delimiter>
Command = CL?<Delimiter>
→
Return = CLm <Delimiter>
Parameter m=3: (Calibration) Source/4: (Calibration) Measure
CMF
Normal Setting
Sets and queries the calibration measurement function
×
○
*
Command = CMFm<Delimiter>
→
Answer = CMFm <Delimiter>
Command = CMF?<Delimiter>
→
Return = CMFm <Delimiter>
Parameter
m=0: DCV/1: DCA/2: Ω
default Value of m=0 (DCV)
CP
Normal Setting
Sets the calibration point
×
○
*
Command = CPm<Delimiter>
→
Answer = CPm <Delimiter>
Parameter
m=0: + FS Calibration/1: + Zero Calibration/2: -FS Calibration
CS
Normal Setting
Sets the measurement value during calibration
×
○
*
Command = CS<Delimiter>
→
Answer = CS <Delimiter>
During measurement in calibration mode, sets the current input
measurement values as the calibration values for the selected function,
range, and scale (+FS/0/-FS ).
CSF
Normal Setting
Sets and queries the calibration source function
×
○
*
Command = CSFm<Delimiter>
→
Answer = CSFm <Delimiter>
Command = CSF?<Delimiter>
→
Return = CSFm <Delimiter>
Parameter m=0: DCV/1: DCA/2:
Ω
default Value of m=0 (DCV)
CW
Normal Setting
Writes the calibration data
×
○
*
Command = CW<Delimiter>
→
Answer = CW, OK<Delimiter> (Normal End)
Writes to EEPROM after calibrating each function and range.
If the power is turned off without executing this command, the values that
were calibrated just before are discarded.
*: Can only be used in calibration mode of setting mode.
8.3 Detailed Description of Commands