Names and Functions of Parts
2-5
IM CA150E
1
2
3
4
5
6
7
8
9
10
11
12
App
2. Names and Functions of Parts
SOURCE: OFF lights when output is off or the protection circuit
has been activated.
ON lights when output is on.
Indicates hold.
Indicates that the 4-20 mA range is selected for the source.
Flashes while communication data is being output.
(When communication of the setting mode is set to printer mode.)
Indicates that 24 V DC is being output for a loop test.
Indicates reading of the memory function.
Lights when the memory function or program sweep function is in use.
(Indicates the memory number.)
Lights when the sweep function is in use.
Lights when the divided output (n/m) function or step sweep
function is in use.
Indicates that contact input is selected during pulse measurement.
Lights or flashes during offset or full scale adjustment in
calibration mode.
Indicates the battery level of the alkaline batteries or NiMH battery.
Indicates that the NiMH battery pack is charging.
Indicates the type of thermocouple.
Lights during an RJ compensation calculation.
Lights when the PT100 or JPT100 standard is selected for the
RTD function.
Lights when IPTS-68 (temperature scale standard) is selected for
temperature source and measurement (TC, RTD).