Specifications
11-5
IM CA150E
1
2
3
4
5
6
7
8
9
10
11
12
App
500 mV
DCV
measurement
TC input
RTD input
20 mA
DCmA
measurement
K
E
J
T
N
L
U
R
S
B
PT100
JPT100
0 to ±500.000 mV
0 to ±20.000 mA
-200.0 to 1372.0 °C
-200.0 to 1000.0 °C
-200.0 to 1200.0 °C
-200.0 to 400.0 °C
-200.0 to 1300.0 °C
-200.0 to 900.0 °C
-200.0 to 400.0 °C
0 to 1768 °C
0 to 1768 °C
600 to 1820 °C
-200.0 to 850.0 °C
-200.0 to 500.0 °C
±(0.035% +0.1 mV)
±(0.04% +6 µA)
±(0.09% + 1.2 °C)
Input resistance: 1000 M
Ω
or greater.
Input resistance:
Approx. 1 M
Ω
Input resistance:
Approx. 20
Ω or less
Measurement current:
Approx. 1 mA
Measurement current:
Approx. 100 µA
Measurement current:
Approx. 10 µA
10 µV
5 V
0 to ±5.0000 V
±(0.035% +1 mV)
0.1 mV
35 V
0 to ±35.000 V
±(0.025% +7 mV)
1 mV
Range
Measurement
Range
Accuracy
Remarks
Resolution
1 µA
500
Ω
Resistance
measurement
0 to 500.00
Ω
±(0.09% +0.15
Ω)
0.01
Ω
5 k
Ω
0 to 5.0000 k
Ω
±(0.09% +1.5
Ω)
0.1
Ω
50 k
Ω
0 to 50.000 k
Ω
±(0.09% +20
Ω)
1
Ω
100 mA
0 to ±100.00 mA
±(0.055% + 40 µA)
10 µA
0.1 °C
1 °C
0.1 °C
––
*4
*4
*6
*5
*6
Measurement during Charging
Accuracy = ± (% of r mV, µA,
Ω
, °C) at 23±5°C
Temperature coefficient: Add the accuracy in normal state ×(1/10)°C.
*6: When the temperature measured with the temperature monitor function is in
the ranges of 0≤T<18°C and 28<T≤40°C,
add the accuracy in the normal state ×(1/10)/°C as a temperature coefficient,
which is obtained from the temperature of the part exceeding 23±5°C.