11-6
IM CA150E
General and Common Specifications
Source unit response time:
300 msec
(time from start of voltage change to when voltage enters accuracy range)
However, the time is 5 msec for 1 V, 10 V, 500 Ω
(excitation current: 1 mA), and RTD (excitation current: 1 mA) ranges
Source unit voltage limiter:
Approximately 32 V
Source unit current limiter:
Approximately 25 mA
Switching of output polarity:
← →
Load condition: 0.01 µF or less (DCV, Ω, TC, RTD, PULSE)
100 µH or less (DCA)
Divided output (n/m) function: Output = setting value × (n/m)
where n = 0 to m and m = 1 to 19
Step sweep function:
The n value is sent automatically
when the n/m function is selected.
The interval time is approximately 5 seconds/
10 seconds
Linear sweep function:
The sweep time is approximately 16 seconds/
32 seconds
Memory function:
1) Setting:
Up to 21 settings for startup such as the function and
range can be saved from the setting screen.
At startup, it is possible to select initial settings or switch output.
2) Data:
100 items of data can be saved
(sets of items such as date, source function, range, source value,
measurement function, range, and measurement value)
Multiple data can be continuously output
(program sweep function)
Loop check function
(current can be measured while supplying 24 V with two wires)
Measurement unit maximum input:Voltage terminal:
42 V DC, Current terminal: 120 mA
Current terminal input protection:
Fuse: 125 mA/250 V
Measurement display update rate:
Approximately once per second
CMRR:
Approximately 120 dB (50/60 Hz)
NMRR:
Approximately 60 dB (50/60 Hz)