9. Calibration Mode
To maintain high accuracy, it is recommended to calibrate the instrument
once a year. This section describes the calibration procedure using
the standard devices recommended in "Standard Device Selection."
(1) Standard Device Selection
Source Function
Function
DCA
DCV
Standard Device
Range
Accuracy
100 mV
1 V
10 V
30 V
20 mA
20 mA SINK
500
Ω
5 k
Ω
50 k
Ω
0.002% + 1 µV
0.002% + 5 µV
0.002% + 0.05 mV
0.002% + 1 µV
0.0035% + 0.5 µA
0.0035% + 0.5 µA
0.002% + 0.01
Ω
0.005% + 0.15
Ω
0.01% + 5
Ω
• Digital multimeter
• Digital multimeter
• DC voltage/current
standard
Ω
Measurement Function
Function
DCA
DCV
Standard Device
Range
Accuracy
500 mV
5 V
35 V
20 mA
100 mA
500
Ω
5 k
Ω
50 k
Ω
0.002% + 5 µV
0.002% + 50 µV
0.0025% + 0.5 mV
0.0025% + 0.4 µA
0.004% + 3 µA
0.0055% + 7.5 m
Ω
0.0055% + 0.75
Ω
0.055% + 1
Ω
• DC voltage/current
standard
Reference resistor
Ω
(2) Required Environment and Conditions for Calibration
Ambient temperature: 23 ±1°C
Relative humidity:
45% to 75%
Warm-up time:
The standard devices must be warmed up for the specified times and
the instrument must be warmed up for at least 2 hours (for calibration).