4.9 Sweep Output Functions
4.9.2 Linear Sweep Function
Output can be varied in a continuous manner as shown in the figure below.
Setting value
Source value indication
Actual output
Approx. 2 sec
Approx. 3 sec
OFF
SOURCE ON indication
Flashing
SOURCE ON/OFF
key operation
SOURCE OFF indication
Flashing
Press
(OFF)
0
Press
(ON)
OFF
OFF
SOURCE ON/OFF
key operation
Press
(ON)
Press
(OFF)
Press
(ON)
When the SOURCE ON
key is pressed while output
is being varied)
1
Set the source value to output (FUNCTION, RANGE, Setting Value).
2
Press the SWEEP SET key twice to select Linear Sweep.
Step Sweep → Linear Sweep
(SWEEP appears.)
3
Use the ▲ ▼ output setting key to set the setting value (upper value).
4
Press the SOURCE ON/OFF key to start output automatically as
shown in the figure above.
5
Press the SOURCE ON/OFF key again to stop the output.
6
Press the SWEEP SET key twice to cancel the sweep output function.
→ Program Sweep → Cancel
Tip
• The default setting for the linear sweep time is 16 seconds.
For details on changing the setting, refer to the next page.
• When output reaches the lower limit, the state is retained for 3 seconds,
and then output is turned off and the first sweep operation ends.
• To change the direction for varying output, press the SOURCE ON/OFF key
during the sweep operation. The direction for varying output changes and
the sweep operation continues. (For example, pressing the SOURCE ON/OFF
key during rising output results in the output switching to a falling direction.)