Communication Function
8-13
IM CA150E
1
2
3
4
5
6
7
8
9
10
11
12
App
SO
Normal Setting
Starts/stops and queries source
○
○
Command = SOm<Delimiter>
→
Answer = SOm <Delimiter>
Command = SO?<Delimiter>
→
Return = SOm <Delimiter>
Parameter
m = 0: Stop/1: Start
SR
Normal Setting
Sets and queries the source range
○
○
Command = SRm<Delimiter>
→
Answer = SRm <Delimiter>
Command = SR?<Delimiter>
→
Return = SRm <Delimiter>
Parameter
[DCV] m = 0: 100 mV/1: 1 V/2: 10 V/3: 30V
[DCA]
(When normal)
m = 0: 20 mA/1: 4-20 mA
(When Calibrating) m = 0: 20 mA/2: 20 mA SINK
[Ω
]
(When normal)
m = 0: 500
Ω
/1: 5 k
Ω
/2: 50 k
Ω
(When Calibrating) m = 0: 500
Ω
(1 mA)/1: 5 k
Ω
(0.1 mA)/
2: 50 k
Ω
(0.05 mA)/
3: 500
Ω
(5 mA)/4: 5 k
Ω
(0.5 mA)
[TC] m = 0: K/1: E/2: J/3: T/4: R/5: B/6: S/ 7: N/8: L/9: U
[RTD] m = 0: PT100/1: JPT100
[PULSE] m = 0: 100 Hz/1: 1000 Hz/2: 10 kHz/3: 50 kHz/4: 1000 CPM
SY
Normal Setting
Switches and queries normal mode and setting mode
○
○
Command = SYm<Delimiter>
→
Answer = SYm <Delimiter>
Command = SY?<Delimiter>
→
Return = SYm <Delimiter>
Parameter
m = 0: Normal Mode/1: Setting Mode
default Value
m = 0 (Normal Mode)
TE
Normal Setting
Sets and queries TC and RTD (source) display
○
×
Command = TEm<Delimiter>
→
Answer = TEm <Delimiter>
Command = TE?<Delimiter>
→
Return = TEm <Delimiter>
Parameter
m = 0: Temperature Value/1: mV Value (Resistance Value)/2: Room
Temperature
default Value
m = 0 (Temperature Value)
TT
Normal Setting
Sets and queries the international temperature standard
×
○
Command = TTm<Delimiter>
→
Answer = TTm <Delimiter>
Command = TT?<Delimiter>
→
Return = TTm <Delimiter>
Parameter m = 0: IPTS-68/1: ITS-90
default Value m = 1 (ITS-90)
8.3 Detailed Description of Commands