95
How to Use the Vivo 65
Vivo 65 users manual
Doc. 006088 En-US U-1b
*
These ventilator settings affect the operating time significantly.
Storing the Internal Battery and the Click-on Battery
Storage longer than one month should be initiated with half-charged bat-
teries in order to maintain maximum capacity. Optimal storage tempera-
ture is 41 to 86°F (5 to 30°C).
External DC
The Vivo 65 can be operated from:
• a 12 V external DC source using the 12/24 V converter
(part no. 004901).
• a 24 V external DC source using the external battery cable
(part no. 004899).
With an external DC source connected, the Vivo 65 will automatically
switch over to the external DC source if the mains power cord is removed
or if the AC power supply fails. The external DC voltage level is shown
under “Others”, “Device Information” in the menu.
Internal Battery
3.5 h
3.5 h
Click-on Battery
8 h
8 h
Do not connect the Vivo 65 to a wheelchair unless the operating manual
for the wheelchair permits this as this can affect the Vivo 65 performance
and consequently result in patient death.
Only use a battery charger compliant to IEC 60601-1 if you are charging a
battery that at the same time is connected to the Vivo 65.
1
Connect the external DC cable
to the Vivo 65. Make sure that it is
fitted correctly.
2
Connect the other end of the cable to the DC source.
P
ARAMETER
E
XAMPLE
1
E
XAMPLE
2