58 How to Use the Vivo 65
Vivo 65 users manual
Doc. 006088 En-US U-1b
PEEP (Positive End Expiratory Pressure)
I
TEM
D
ESCRIPTION
Definition
The PEEP setting is used to define the air-
way pressure at the end of the expiratory
phase.
Modes
PSV, PSV(TgV), PCV, PCV(TgV), PCV(A),
PCV(A+TgV), PCV-SIMV, VCV, VCV(A),
VCV-SIMV
Setting min
Off, 2 cmH
2
O (leakage circuit, circuit with
external/integrated active exhalation valve,
single limb exhalation valve for Pediatric
VCV)
Setting max
20 cmH
2
O (Adult, Paediatric), Pressure -
2 cmH
2
O or Min Pressure -2 cmH
2
O.
Setting resolution
0.5 below 10 cmH
2
O,
1.0 above 10 cmH
2
O
Display