69
How to Use the Vivo 65
Vivo 65 users manual
Doc. 006088 En-US U-1b
Expiratory Trigger (Exp. Trigger)
I
TEM
D
ESCRIPTION
Definition
The Expiratory Trigger setting defines the
moment when the ventilator will cycle from
the inspiratory to the expiratory phase.
In PCV-SIMV and VCV-SIMV modes, this
setting is applicable to the pressure support
breaths that are triggered by the patient.
Modes
PSV, PSV(TgV), PCV-SIMV, VCV-SIMV
Setting min
1 (10% decrease of peak flow)
Setting max
9 (90% decrease of peak flow)
Setting resolution
1
(setting 1 cycles off early and 9 cycles off
late)
Display