146 Alarms
Vivo 65 users manual
Doc. 006088 En-US U-1b
Exhalation Valve Control Error Alarm
I
TEM
D
ESCRIPTION
Definition
An Exhalation Valve Control Error alarm will
be given when the Vivo 65 fails to control the
internal /external exhalation valve.
Priority
High
Possible cause
• Malfunction in the exhalation valve.
• Internal error of the ventilator.
Indication
The alarm is given audibly with a tone and
visible by the red alarm LED and a display
message.