193
Technical Specifications
Vivo 65 users manual
Doc. 006088 En-US U-1b
Guidance and Manufacturer's Declaration – Electromag-
netic Immunity
The Vivo 65 is intended for use in the electromagnetic environment spec-
ified below. The customer or the user of the Vivo 65 should assure that it
is used in such an environment.
I
MMUNITY
T
EST
C
OMPLIANCE
L
EVEL
E
LECTROMAGNETIC
E
NVIRONMENT
-
G
UIDANCE
Electrostatic discharge
(ESD) IEC 61000-4-2
±8 kV contact
±15 kV air
The relative humidity
should be at least 5 %.
Electrical fast transient/
burst IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality
should be that of a typi-
cal commercial, hospital
and residential environ-
ment.
Surge IEC 61000-4-5
±1 kV line(s) to
line(s)
Mains power quality
should be that of a typi-
cal commercial, hospital
and residential environ-
ment.
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
30 A/m
Power frequency mag-
netic fields should be at
levels characteristic of a
typical location in a typ-
ical commercial, hospi-
tal and residential
environment.
Voltage dips, short
interruptions and volt-
age variations on power
supply input lines
IEC 61000-4-11
0 % UT, 0.5 cycle (mul-
tiple phase analysis);
0 % UT, 1 cycle;
70 % UT, 25/30 cycles
(50/60 Hz);
0 % UT, 250/300 cycles
(50/60 Hz).
Vivo 65 runs on internal
battery during voltage
dips, short interruptions
and voltage variations
on power supply input
lines.
U
T is the mains voltage prior to application of the test level.