178 Technical Specifications
Vivo 65 users manual
Doc. 006088 En-US U-1b
PEEP
Off, 2 cmH
2
O (leakage
circuit, circuit with active
exhalation valve, (external/
integrated), single limb
exhalation valve for Pediat-
ric VCV) to 20 cmH
2
O
(Adult),
to 20 cmH
2
O (Pediatric),
Insp. Pressure -2 cmH
2
O
or
Min Pressure -2 cmH
2
O.
Tolerance:
±0.5 cmH
2
O or ±5%,
whichever is greatest.
0.5 below 10 cmH
2
O,
1.0 above 10 cmH
2
O
Breath Rate
4 to 40 bpm (breaths per
minute) (Adult),
6 to 60 bpm (Pediatric).
0 to 40/60 bpm (MPV
breath mode)
Tolerance: ±2%
1 bpm
SIMV Rate
4 to 40 bpm (breaths per
minute) (Adult),
6 to 60 bpm (Pediatric).
Tolerance: ±2%
1 bpm
Inspiratory
Time
0.3 to 5 s (Adult),
0.3 to 2 s (Pediatric).
Tolerance: ±0.1 s
0.1 s
Backup
Inspiratory
Time
0.3 to 5 s (Adult),
0.3 to 2 s (Pediatric).
0.1 s
S
ETTING
/V
ALUE
R
ANGE
/P
ERFORMANCE
R