141
Alarms
Vivo 65 users manual
Doc. 006088 En-US U-1b
Low Last Power Source Alarm
I
TEM
D
ESCRIPTION
Definition
A Low Last Power Source alarm will be given
when the last battery source (internal battery
or click-on battery) has 15 minutes of operat-
ing time left with current settings.
Priority
Medium
Ventilator action
The Vivo 65 will continue treatment with the
same settings.
Indication
The alarm is given audibly with a tone and
visibly by the yellow alarm LED and a display
message.