74 How to Use the Vivo 65
Vivo 65 users manual
Doc. 006088 En-US U-1b
Max Pressure
I
TEM
D
ESCRIPTION
Definition
The Max Pressure setting is only used when
Target Volume is activated. Max Pressure
defines the upper pressure limit up to where
the Vivo 65 can increase the pressure to
reach the set Target Volume. If Target Vol-
ume is not reached at Max Pressure, the
Vivo 65 will continue to ventilate at this Max
Pressure setting.
Modes
PSV(TgV), PCV(TgV), PCV(A+TgV)
Setting min
Min Pressure
Setting max
50 cmH
2
O or Max Pressure (Adult)
35 cmH
2
O or Max Pressure (Pediatric)
Setting resolution
0.5 below 10 cmH
2
O,
1.0 above 10 cmH
2
O
Display