143
Alarms
Vivo 65 users manual
Doc. 006088 En-US U-1b
Critical Low Last Power Source Alarm
I
TEM
D
ESCRIPTION
Definition
A Critical Low Last Power Source alarm will
be given when the last battery source (internal
battery or click-on battery) has at least 5 min-
utes of operating time left with current set-
tings.
Priority
High
Ventilator action
The Vivo 65 will continue treatment with the
same settings.
Indication
The alarm is given audibly with a tone and
visibly by the red alarm LED and a display
message.