66 How to Use the Vivo 65
Vivo 65 users manual
Doc. 006088 En-US U-1b
Rise Time
I
TEM
D
ESCRIPTION
Definition
The Rise Time setting controls the speed of
the pressure/volume increase from start of
inspiration to set inspiratory pressure/tidal
volume.
A low setting will give a faster pressure/vol-
ume increase and therefore a longer plateau
at the set inspiratory pressure/tidal volume.
A high setting will give a slow increase and
therefore a shorter plateau.
In the VCV-SIMV mode, two different Rise
Time settings are made, one for mandatory
ventilator-controlled breaths and one for
additional pressure support breaths trig-
gered by the patient.
Modes
PSV, PSV(TgV), PCV, PCV(TgV), PCV(A),
PCV(A+TgV), PCV-SIMV, PCV-MPV,
VCV, VCV(A), VCV-SIMV, VCV-MPV
Setting min
1 (PSV & PCV, PCV-SIMV, PCV-MPV,
VCV-SIMV)
50% of the inspiration time (Min. 0.3 s)
(VCV, VCV-SIMV, VCV-MPV)
Setting max
9 (PSV & PCV, PCV-SIMV, PCV-MPV,
VCV-SIMV)
90% of the inspiration time, Off (=100%)
(VCV, VCV-SIMV, VCV-MPV)
Setting resolution
1 (PSV & PCV),
10% (VCV)
Display