123
Alarms
Vivo 65 users manual
Doc. 006088 En-US U-1b
High Breath Rate Alarm
I
TEM
D
ESCRIPTION
Definition
A High Breath Rate alarm will be given when the
delivered total breath rate exceeds the High
Breath Rate alarm limit for 15 ±0.5 seconds.
Priority
Medium
Possible cause
• Mismatch between the Breath Rate setting and
the alarm setting.
• Increased Breath Rate.
• Too sensitive setting of the inspiratory trigger
setting.
Setting min
10 bpm
Setting max
70 bpm, Off (Adult),
99 bpm, Off (Pediatric)
Setting resolution
1 bpm
Ventilator action
The Vivo 65 will continue treatment with the
same settings.
Indication
The alarm is given audibly with a tone and visibly
by the yellow alarm LED and a display message.