15. Appendices
Green Smart User Manual 157
EN
G
LISH
15.2 X-ray Dose Data
15.2.1 DAP (Dose Area Product)
The X-ray dose data is extracted from the X-ray Dose Test Report for the
Green Smart
.
X-ray Dose Test Report for the
Green Smart
maintains dosimetric evaluation that the
VATECH
dental diagnostic system meets all requirements specified in the IEC Collateral
Standard. To limit unnecessary exposure to the patient, operator or other staff, the
Green
Smart
is designed to comply with IEC 60601-1-3 Part 1 General Requirements for Safety.
Test Hardware
Brand Name (Model) Green Smart (Model: PHT-35LHS)
Sensor Type
PANO & CBCT: Xmaru1404CF-Plus
CEPH: Xmaru2602CF
X-ray Generator
DG-07E22T2
Tube
D-052SB
DAP (Dose Area Product) is a quantity used in assessing the radiation risk from diagnostic
X-ray examination procedures. It is defined as the absorbed dose multiplied by the area
irradiated, expressed in gray square centimeters (mGy·cm
2
). Despite the limitation, DAP is
the best way to predict effective dose value and currently the most convenient method for
patient doses monitoring.
PHE (Public Health England) recommends that any national reference dose that
achievable dose (DAP) value of 250 [mGy·cm2] for a clinical protocol for a standard
male patient.
Standard
National
Deviations
Terminology
Permissive Range
PHE (GBR)
DAP
(Pano, CEPH
and CBCT)
PANO
: ≤ 93 mGy
·cm
2
(Adult) / 67 mGy·cm
2
(Child)
CEPH
: ≤ 40 mGy
·cm² (Adult, Lateral) / 25
mGy·cm² (Child, Lateral)
CBCT
: ≤ 250 mGy
·cm
2
(Basis of single
implant)
AERB (IND)
Dose
All dose values must be within ± 20 % of the
SPECIFIED values
Summary of Contents for GREEN SMART
Page 1: ...User Manual Version 1 23 English...
Page 2: ......
Page 3: ......
Page 4: ......
Page 52: ...5 Imaging Software Overview 44 Green Smart User Manual Left blank intentionally...
Page 90: ...7 Acquiring PANO Images 82 Green Smart User Manual Left blank intentionally...
Page 110: ...8 Acquiring CEPH Images Optional 102 Green Smart User Manual Left blank intentionally...
Page 126: ...9 Acquiring CBCT Images 118 Green Smart User Manual Left blank intentionally...
Page 134: ...10 Acquiring 3D MODEL Scan Images 126 Green Smart User Manual Left blank intentionally...
Page 142: ...12 Cleaning and Maintenance 134 Green Smart User Manual Left blank intentionally...
Page 144: ...13 Disposing of the Equipment 136 Green Smart User Manual Left blank intentionally...
Page 154: ...14 Technical Specifications 146 Green Smart User Manual Left blank intentionally...
Page 185: ...15 Appendices Green Smart User Manual 177 ENGLISH...
Page 200: ......