8. Acquiring CEPH Images (Optional)
Green Smart User Manual 91
EN
G
LISH
6.
Align the Ear Rods to the patient’s ears properly so that the head does not move during
the operation. Moreover, align the Nasal Positioner with the patient’s nation by adjusting its
height.
7.
Align horizontally, so the patient's Frankfurt Line is parallel to the floor.
8.
Direct the patient to swallow first before closing the mouth and to remain in his/her current
position until image acquisition is completed.
9.
Click the
READY
button. The x-ray exposure has not started yet.
10.
Now go to
8.4 X-ray Exposure
to start the exposure.