7. Acquiring PANO Images
Green Smart User Manual 69
EN
G
LISH
Laser Beam Aligning
Be careful not to shine the laser beam directly into the person's eyes.
Doing so may result in vision loss.
If the laser beam is not correctly positioned, there may be distortion,
causing the image to be enlarged or reduced, or ghost shadows may
occur and lower the image quality. Be sure to align the laser beam
properly.
1.
Align the Vertical Beam with the center of the face (Mid-sagittal Line). (It’s to prevent
horizontal expansion of the image.)
2.
Align the Horizontal Beam in a straight line to the Frankfurt Line on the patient's face. Use
the
Horizontal Beam
button on the control panel to position it. Make sure that the
Horizontal Beam is aligned to the patient's face horizontally.
3.
Direct the patient to smile and align the Canine Teeth Beam to the center of the canines.
Use the Canine Teeth Beam Lever to adjust the position of the beam.
Summary of Contents for GREEN SMART
Page 1: ...User Manual Version 1 23 English...
Page 2: ......
Page 3: ......
Page 4: ......
Page 52: ...5 Imaging Software Overview 44 Green Smart User Manual Left blank intentionally...
Page 90: ...7 Acquiring PANO Images 82 Green Smart User Manual Left blank intentionally...
Page 110: ...8 Acquiring CEPH Images Optional 102 Green Smart User Manual Left blank intentionally...
Page 126: ...9 Acquiring CBCT Images 118 Green Smart User Manual Left blank intentionally...
Page 134: ...10 Acquiring 3D MODEL Scan Images 126 Green Smart User Manual Left blank intentionally...
Page 142: ...12 Cleaning and Maintenance 134 Green Smart User Manual Left blank intentionally...
Page 144: ...13 Disposing of the Equipment 136 Green Smart User Manual Left blank intentionally...
Page 154: ...14 Technical Specifications 146 Green Smart User Manual Left blank intentionally...
Page 185: ...15 Appendices Green Smart User Manual 177 ENGLISH...
Page 200: ......