9. Acquiring CBCT Images
Green Smart User Manual 115
EN
G
LISH
Laser Beam Aligning
Be careful not to shine the Laser Beam directly into the person's eyes.
Doing so may result in vision loss.
If the Laser Beam is not correctly positioned, there may be distortion,
causing the image to be enlarged or reduced, or ghost shadows may
occur and lower the image quality. Be sure to align Laser Beam
properly.
This is a sample illustration for reference only. Actual FOV may vary
from the image as shown above.
1.
Align the Vertical Beam with the center of the face (Mid-sagittal Line). (It’s to prevent
horizontal expansion of the image.)
Finishing Patient Positioning
1.
After checking the positions of the patient and the Laser Beam, click the
Temple
Supports OPEN/CLOSE
button on the control panel to prevent the patient's head from
moving.
Make sure that the Temple Supports are in CLOSE position before
clicking the
READY
button.
2.
Click the
READY
button. The X-ray exposure has not started yet.
3.
Now go to
9.4 X-ray Exposure
to start the exposure.
Summary of Contents for GREEN SMART
Page 1: ...User Manual Version 1 23 English...
Page 2: ......
Page 3: ......
Page 4: ......
Page 52: ...5 Imaging Software Overview 44 Green Smart User Manual Left blank intentionally...
Page 90: ...7 Acquiring PANO Images 82 Green Smart User Manual Left blank intentionally...
Page 110: ...8 Acquiring CEPH Images Optional 102 Green Smart User Manual Left blank intentionally...
Page 126: ...9 Acquiring CBCT Images 118 Green Smart User Manual Left blank intentionally...
Page 134: ...10 Acquiring 3D MODEL Scan Images 126 Green Smart User Manual Left blank intentionally...
Page 142: ...12 Cleaning and Maintenance 134 Green Smart User Manual Left blank intentionally...
Page 144: ...13 Disposing of the Equipment 136 Green Smart User Manual Left blank intentionally...
Page 154: ...14 Technical Specifications 146 Green Smart User Manual Left blank intentionally...
Page 185: ...15 Appendices Green Smart User Manual 177 ENGLISH...
Page 200: ......